The load transfer effect of monopolar and bipolar radial head prosthesis designs on capitellar cartilage: A comparative biomechanical cadaver study

Volume 56 · Issue 1 · January 2022 Publish Date: February 11, 2022
Full Text PDF
DOI
Mehmet Ozan Aşık
Clinic of Orthopaedics and Traumatology, Özel Adatıp Hospital, İstanbul, Turkey
Mehmet Akdemir
Clinic of Orthopaedics and Traumatology, Özel Akademi Hospital, Konya, Turkey
Necmettin Turgut
Department of Orthopedics and Traumatology, Isparta City Hospital, Isparta, Turkey
Abdullah Meriç Ünal
Clinic of Orthopaedics and Traumatology, Özel Meddem Hospital, Isparta, Turkey
Mustafa Özkan
Department of Orthopaedics and Traumatology, Dokuz Eylül University, School of Medicine, İzmir, Turkey
Ahmet Ekin
Clinic of Orthopaedics and Traumatology, Medical Park Hospital, İzmir, Turkey
Aşık, M. O., Akdemir, M., Turgut, N., Ünal, A. M., Özkan, M., & Ekin, A. (2022). The load transfer effect of monopolar and bipolar radial head prosthesis designs on capitellar cartilage: A comparative biomechanical cadaver study. Acta Orthopaedica Et Traumatologica Turcica, 56(1), 58–63. https://doi.org/10.5152/j.aott.2021.20410
Full Text Full Text PDF

Abstract

Objective: The aim of this study was to determine whether bipolar radial head arthroplasty may transfer less load to the capitellum than monopolar radial head arthroplasty and native radial heads.

Methods: Six human elbow joints were obtained from six fresh frozen cadavers (3 males, 3 females; mean age = 78 years, age range = 66-80). None of the elbow joints had a previous osseous injury, a chondral defect, osteoarthritis, or instability. In group 1, load transfer of the native radial head was measured in each specimen under 100 N of compression force using a custom-made load cell at 0°, 30°, 60°, 90°, and 120° flexion of the elbow in supination, neutral rotation, and pronation, respectively. After excision of the radial heads, the same testing protocol was first applied for monopolar radial head arthroplasty (Group 2) and then for bipolar radial head arthroplasty (Group 3).

Results: The mean load transfer on the capitellum was significantly higher in each forearm rotation and all angles of the elbow flexion in the arthroplasty groups than the native radial head group. Mean load transfer values of bipolar prostheses were between the values of native radial heads and monopolar prostheses in all positions. Bipolar prostheses showed similar load transfer characteristics compared to those of the native radial head in supination at 60° flexion; in neutral rotation at 0°, 30°, 60°, and 120° flexion; and in pronation at 90° and 120° flexion.

Conclusion: The results of this study have revealed that bipolar radial head arthroplasty transfers similar loading as the native radial head on the capitellum in certain forearm positions and at elbow flexion angles. No significant differences could be found between load transfer values of bipolar head design and monopolar head design except in the pronation at full extension.

Cite this article as: Aşık MO, Akdemir M, Turgut N, et al. The load transfer effect of monopolar and bipolar radial head prosthesis designs on capitellar cartilage: A comparative biomechanical cadaver study. Acta Orthop Traumatol Turc 2022; 56(1):59–64.

Similar Articles

Article Info
Published In
Journal Acta Orthopaedica et Traumatologica Turcica
Volume / Issue Volume 56 · Issue 1 · January 2022
Pages 58-63
History
Published Online February 11, 2022
Copyright
Affiliations
1
Mehmet Ozan Aşık
Clinic of Orthopaedics and Traumatology, Özel Adatıp Hospital, İstanbul, Turkey
2
Mehmet Akdemir
Clinic of Orthopaedics and Traumatology, Özel Akademi Hospital, Konya, Turkey
3
Necmettin Turgut
Department of Orthopedics and Traumatology, Isparta City Hospital, Isparta, Turkey
4
Abdullah Meriç Ünal
Clinic of Orthopaedics and Traumatology, Özel Meddem Hospital, Isparta, Turkey
5
Mustafa Özkan
Department of Orthopaedics and Traumatology, Dokuz Eylül University, School of Medicine, İzmir, Turkey
6
Ahmet Ekin
Clinic of Orthopaedics and Traumatology, Medical Park Hospital, İzmir, Turkey
Cite this Article
Aşık, M. O., Akdemir, M., Turgut, N., Ünal, A. M., Özkan, M., & Ekin, A. (2022). The load transfer effect of monopolar and bipolar radial head prosthesis designs on capitellar cartilage: A comparative biomechanical cadaver study. Acta Orthopaedica Et Traumatologica Turcica, 56(1), 58–63. https://doi.org/10.5152/j.aott.2021.20410
Outlines