Acta Orthopaedica et Traumatologica Turcica

Novel ultrasound assisted suture anchor system using the BoneWelding® technology yields a comparable primary stability in osteopenic and healthy human humeri as a benchmark anchor

AOTT 2018; 52: 127-133
DOI: 10.1016/j.aott.2017.11.009
Read: 1444 Downloads: 658 Published: 06 February 2020
Abstract

Introduction
The aim of this biomechanical study was to evaluate the primary stability of the SportWelding® Sombrero 3.6 mm suture anchor system in osteopenic and healthy cadaveric humeri.
Methods
The Sombrero® and BioCorkscrew® anchors were deployed in 8 osteopenic and 4 healthy cadaver humeri after the bone mineral density (BMD) measurements of the 32 specimens. Both anchors were loaded with a USP Nr. 2 FiberWire® suture. An established cyclic testing protocol was performed. The maximum failure load (Fmax), the system displacement and the modes of failure were recorded.
Results
The Fmax and system displacement of the Sombrero® in osteopenic and healthy humeri was equivalent to the Bio-Corkscrew® benchmark anchor; there were no significant differences in the maximum failure loads and system displacement values. Only anchor and suture dislocations were observed; suture ruptures did not occur.
Conclusion
This study shows that the Sombrero® yields similar maximum failure loads and system displacement values as the established Bio-Corkscrew® benchmark anchor. The primary stability of the Sombrero® and Bio-Corkscrew® seems to be independent of the bone mineral quality. This relatively small-sized polymer anchor is independent of the BMD and may be an alternative to established suture anchors in rotator cuff repair.
ER -

Keywords

  • Biomechanics
  • Ultrasound
  • Suture anchor
  • Rotator cuff
  • Osteoporosis
  • Bone welding
  • Bone density

 

 

 

Files
ISSN 1017-995X EISSN 2589-1294