Abstract
Introduction
The aim of this biomechanical study was to evaluate the primary stability of the SportWelding® Sombrero 3.6 mm suture anchor system in osteopenic and healthy cadaveric humeri.
Methods
The Sombrero® and BioCorkscrew® anchors were deployed in 8 osteopenic and 4 healthy cadaver humeri after the bone mineral density (BMD) measurements of the 32 specimens. Both anchors were loaded with a USP Nr. 2 FiberWire® suture. An established cyclic testing protocol was performed. The maximum failure load (Fmax), the system displacement and the modes of failure were recorded.
Results
The Fmax and system displacement of the Sombrero® in osteopenic and healthy humeri was equivalent to the Bio-Corkscrew® benchmark anchor; there were no significant differences in the maximum failure loads and system displacement values. Only anchor and suture dislocations were observed; suture ruptures did not occur.
Conclusion
This study shows that the Sombrero® yields similar maximum failure loads and system displacement values as the established Bio-Corkscrew® benchmark anchor. The primary stability of the Sombrero® and Bio-Corkscrew® seems to be independent of the bone mineral quality. This relatively small-sized polymer anchor is independent of the BMD and may be an alternative to established suture anchors in rotator cuff repair.
ER -
Keywords
- Biomechanics
- Ultrasound
- Suture anchor
- Rotator cuff
- Osteoporosis
- Bone welding
- Bone density